Analisis Keakurasian Sistem Akuisisi Data Pada
Rancang Bangun Kalibrator Sphygmomanometer
Dublin Core
Title
PROSIDING-NASIONAL-KEBIDANAN-2020 POLTEKES KEMENKES CALL FOR PAPER VOL.2 NO 1
Analisis Keakurasian Sistem Akuisisi Data Pada
Rancang Bangun Kalibrator Sphygmomanometer
Analisis Keakurasian Sistem Akuisisi Data Pada
Rancang Bangun Kalibrator Sphygmomanometer
Subject
Calibration; sphygmomanometer; calibrator;data acquisition
Description
Abstract— This study aims to analyze the accuracy of the data acquisition system in the design of the sphygmomanometer calibrator.
The contribution of this research to the tool will be complemented by sending measurement data via the internet so that it has a further
data transmission range. Thus, after the calibration, the data can be directly sent and processed by the engineering manager with
automatic data calculation. This tool is made to facilitate the work of health workers, especially calibration officers to perform calibrations
and perform data analysis processing. The sensors used in this tool are the MPX5100 pressure sensor to detect air pressure, the
Microcontroller IC, a program to retrieve and send data by Bluetooth to a smartphone. This tool uses a smartphone as a means of
correcting calibration data, inputting other information data and sending working paper data to the internet. From the data obtained,
the% error rate for the total DPM Module against the standard DPM from the increasing average measurement is 0.02% while the
measurement decreases by 0.04%, so that the DPM module has measurement results close to the correct value against the standard DPM.
The results of data loss analysis are 0% with different delivery speed levels according to the spesifikcation of network providers, so it can be
concluded that the DPM module design has good data acquisition accuracy against the standard DPM. Research development can be
done by adding sensors for negative pressure measurement and developing a system for applications on smartphones.
Keywords : Calibration; sphygmomanometer; calibrator;data acquisition
Abstrak— Untuk menjaga keakuratan hasil pengukuran Sphygmomanometer diperlukan kalibrasi secara berkala. Alat untuk
kalibrasi Sphygmomanometer adalah Digital Pressure Meter (DPM). Penelitian ini bertujuan untuk menganalisis keakurasian sistem
akuisisi data pada rancang bangun kalibrator sphygmomanometer. Kontribusi penelitian ini pada alat akan dilengkapi dengan
pengiriman data pengukuran melalui internet sehingga memiliki jangkauan pengiriman data yang lebih jauh. Dengan demikian setelah
dilakukan kalibrasi, data dapat langsung dikirim dan diolah oleh manager teknik dengan perhitungan data secara otomatis. Alat ini
dibuat untuk memudahkan pekerjaan tenaga kesehatan, khususnya petugas kalibrasi untuk melakukan kalibrasi dan melakukan
pengolahan analisis data. Sensor yang digunakan pada alat ini yaitu sensor tekanan MPX5100 untuk mendeteksi tekanan udara, IC
Mikrokontroler, program untuk mengambil dan pengiriman data dengan bluetooth ke smartphone. Alat ini menggunakan smartphone
sebagai sarana koreksi data kalibrasi, input data keterangan lainya dan sebagai pengirim data kertas kerja ke internet. Dari data yang
dapatkan Tingkat % error total DPM Modul terhadap DPM standar dari rata rata pengukuran naik adalah 0,02% sedangkan pada
pengukuran turun sebesar 0,04%, sehingga DPM modul memiliki hasil pengukuran mendekati nilai benar terhadap DPM standar. Hasil
analisis loss data sebesar 0% dengan tingkat kecepatan pengiriman yang berbeda sesuai provider jaringan yang tersedia, sehingga dapat
disimpulkan rancang bangun DPM modul memiliki keakurasian akuisisi data yang baik terhadap DPM standar. Pengembangan
penelitian dapat dilakukan dengan melakukan penambahan sensor untuk pengukuran tekanan negative dan pengembangan sistem pada
aplikasi di smartphone.
Kata kunci : Kalibrasi; sphygmomanometer; kalibrator; akuisisi data
The contribution of this research to the tool will be complemented by sending measurement data via the internet so that it has a further
data transmission range. Thus, after the calibration, the data can be directly sent and processed by the engineering manager with
automatic data calculation. This tool is made to facilitate the work of health workers, especially calibration officers to perform calibrations
and perform data analysis processing. The sensors used in this tool are the MPX5100 pressure sensor to detect air pressure, the
Microcontroller IC, a program to retrieve and send data by Bluetooth to a smartphone. This tool uses a smartphone as a means of
correcting calibration data, inputting other information data and sending working paper data to the internet. From the data obtained,
the% error rate for the total DPM Module against the standard DPM from the increasing average measurement is 0.02% while the
measurement decreases by 0.04%, so that the DPM module has measurement results close to the correct value against the standard DPM.
The results of data loss analysis are 0% with different delivery speed levels according to the spesifikcation of network providers, so it can be
concluded that the DPM module design has good data acquisition accuracy against the standard DPM. Research development can be
done by adding sensors for negative pressure measurement and developing a system for applications on smartphones.
Keywords : Calibration; sphygmomanometer; calibrator;data acquisition
Abstrak— Untuk menjaga keakuratan hasil pengukuran Sphygmomanometer diperlukan kalibrasi secara berkala. Alat untuk
kalibrasi Sphygmomanometer adalah Digital Pressure Meter (DPM). Penelitian ini bertujuan untuk menganalisis keakurasian sistem
akuisisi data pada rancang bangun kalibrator sphygmomanometer. Kontribusi penelitian ini pada alat akan dilengkapi dengan
pengiriman data pengukuran melalui internet sehingga memiliki jangkauan pengiriman data yang lebih jauh. Dengan demikian setelah
dilakukan kalibrasi, data dapat langsung dikirim dan diolah oleh manager teknik dengan perhitungan data secara otomatis. Alat ini
dibuat untuk memudahkan pekerjaan tenaga kesehatan, khususnya petugas kalibrasi untuk melakukan kalibrasi dan melakukan
pengolahan analisis data. Sensor yang digunakan pada alat ini yaitu sensor tekanan MPX5100 untuk mendeteksi tekanan udara, IC
Mikrokontroler, program untuk mengambil dan pengiriman data dengan bluetooth ke smartphone. Alat ini menggunakan smartphone
sebagai sarana koreksi data kalibrasi, input data keterangan lainya dan sebagai pengirim data kertas kerja ke internet. Dari data yang
dapatkan Tingkat % error total DPM Modul terhadap DPM standar dari rata rata pengukuran naik adalah 0,02% sedangkan pada
pengukuran turun sebesar 0,04%, sehingga DPM modul memiliki hasil pengukuran mendekati nilai benar terhadap DPM standar. Hasil
analisis loss data sebesar 0% dengan tingkat kecepatan pengiriman yang berbeda sesuai provider jaringan yang tersedia, sehingga dapat
disimpulkan rancang bangun DPM modul memiliki keakurasian akuisisi data yang baik terhadap DPM standar. Pengembangan
penelitian dapat dilakukan dengan melakukan penambahan sensor untuk pengukuran tekanan negative dan pengembangan sistem pada
aplikasi di smartphone.
Kata kunci : Kalibrasi; sphygmomanometer; kalibrator; akuisisi data
Creator
Adi Pramudono#
, Andjar Pudji , Tribowo Indrato
, Andjar Pudji , Tribowo Indrato
Source
semnas.poltekkesdepkes-sby.ac.id 1
Date
Surabaya, 28 Nopember 2020
Contributor
PERI IRAWAN
Format
PDF
Language
INDONESIA
Type
TEXT
Files
Citation
Adi Pramudono#
, Andjar Pudji , Tribowo Indrato, “PROSIDING-NASIONAL-KEBIDANAN-2020 POLTEKES KEMENKES CALL FOR PAPER VOL.2 NO 1
Analisis Keakurasian Sistem Akuisisi Data Pada
Rancang Bangun Kalibrator Sphygmomanometer,” Repository Horizon University Indonesia, accessed April 11, 2025, https://repository.horizon.ac.id/items/show/113.
Analisis Keakurasian Sistem Akuisisi Data Pada
Rancang Bangun Kalibrator Sphygmomanometer,” Repository Horizon University Indonesia, accessed April 11, 2025, https://repository.horizon.ac.id/items/show/113.