Analisis Filter Digital ( Filter IIR pada ECG Pada
Rancang Bangun ECG Lead I, II, Dan III)
Dublin Core
Title
PROSIDING-NASIONAL-KEBIDANAN-2020 POLTEKES KEMENKES CALL FOR PAPER VOL.2 NO 1
Analisis Filter Digital ( Filter IIR pada ECG Pada
Rancang Bangun ECG Lead I, II, Dan III)
Analisis Filter Digital ( Filter IIR pada ECG Pada
Rancang Bangun ECG Lead I, II, Dan III)
Subject
Keyword—ECG 3 LEAD; Heart Monitoring; Arduino Microcontroller
Description
Abstract— Measurement of bio signals such as electrocardiograph, has an interpretation of noise from other signals. This noise can
interfere with the measurement and make the measurement inaccurate. In previous studies still using analog filters that were applied, the
aim of this study was to create a 3-Lead Electrocardiogram module with Arduino-based Digital Filters. By using a digital filter. The
contribution of this research is the use of digital filters to remove noise in the electrocardiograph signal. This can be achieved by evaluating
the use of filters on the ECG signal. This research uses Infinite Impulse Filter digital filters such as Butterworth, Chebyshev I, Chebyshev
II, and Elliptic in orders 2, 4, 6, 8, and 10. The research was conducted by providing input from the Fuction Generator on Arduino which
has been applied a digital filter with Frequency 0.5Hz – 100Hz cut-off. Apparatus compared to factory electrocardiograph. The filter
measurement uses a data input of 460. Butterworth filter has the greatest emphasis on the order of 8, the frequency of 0.5Hz produces an
emphasis of -5.74298158 dB and a frequency of 100Hz produces an emphasis of -5.93529424 dB. Chebyshev I filter the greatest emphasis
on the order 6 frequency of 0.5Hz produces a suppression of -3.27104076 dB and on the order of 8 the frequency of 100Hz produces a
suppression of -5.08730424 dB. Chebyshev II filter with the greatest emphasis on the order 8, the frequency of 0.5Hz produces a
suppression of -44.66011104 dB and a frequency of 80Hz produces a suppression of -37.33653957 dB. Elliptic filters with the greatest
emphasis on order 6, the frequency of 0.5Hz produces a suppression of -1.55429354 dB and a frequency of 100Hz on the order of 8
produces a suppression of -2.2849115 dB. The results showed that according to the cut-off frequency is the 6th order Butterworth filter
which is suitable for the application of an electrocardiograph signal filter because it has a bandwidth that suppresses signals outside the
cut-off frequency. The results of this research can be implemented in the 3-Lead EKG module in order to eliminate noise or interference
when tapping the ECG signal.
Keyword—ECG 3 LEAD; Heart Monitoring; Arduino Microcontroller
Abstrak— Pengukuran bio sinyal seperti Elektrokardiograf, memiliki interpretasi noise dari sinyal lain. Noise tersebut
dapat menggangu dalam pengukuran dan membuat pengukuran menjadi tidak akurat. Dalam penelitian terdahulu masih
menggunakan filter analog yang d aplikasikan, maka tujuan dari penelitian ini adalah dibuatnya modul Elektrokardiogram
3-Lead dengan Filter Digital Berbasis Arduino. Dengan mengunakan filter digital. Konstribusi penelitian ini adalah
penggunaan filter digital untuk menghilangkan noise pada sinyal elektrokardiograf. Ini dapat dicapaidengan mengevaluasi
penggunaan filter pada siyal ECG. Penelitian ini menggunakan filter digital Infinite Impulse Filter seperti Butterworth,
Chebyshev I, Chebyshev II, dan Elliptic pada orde 2, 4, 6, 8, dan 10. Penelitian dilakukan dengan memberikan input dari
Fuction Generator pada Arduino yang telah di aplikasikan filter digital dengan Frekuensi cut-off 0.5Hz–100Hz. Alat
dibandingkan dengan Elektrokardiograf pabrik. Pengukuran filter menggunakan input data sebesar 460. Filter
Butterworth penekanan terbesar pada orde 8 frekuensi 0.5Hz menghasilkan penekanan -5,74298158 dB dan frekuensi
100Hz menghasilkan penekanan -5,93529424 dB. Filter Chebyshev I penekanan terbesar pada orde 6 frekuensi 0.5Hz
menghasilkan penekanan -3,27104076 dB dan pada orde 8 frekuensi 100Hz menghasilkan penekanan -5,08730424 dB. Filter
Chebyshev II penekanan terbesar pada orde 8 frekuensi 0.5Hz menghasilkan penekanan -44,66011104 dB dan frekuensi
80Hz menghasilkan penekasan -37,3653957 dB. Filter Elliptic penekanan terbesar pada orde 6 frekuensi 0.5Hz
menghasilkan penekanan -1,55429354 dB dan frekuensi 100Hz pada orde 8 menghasilkan penekanan -2,2849115 dB. Hasil
penelitian menunjukkan bahwa yang sesuai dengan frekuensi cut-off yaitu filter Butterworth orde 6 yang sesuai untuk
pengaplikasian filter sinyal Elektrokardiograf karena memiliki bandwidth yang menekan sinyal diluar frekuensi cut-off.
Hasil penelitian ini dapat diimplementasikan pada modul EKG 3-Lead agar menghilangkan noise atau gangguan pada saat
penyadapan sinyal ECG.
Kata Kunci—ECG 6 LEAD; Heart Monitoring; Arduino Microcontroller
interfere with the measurement and make the measurement inaccurate. In previous studies still using analog filters that were applied, the
aim of this study was to create a 3-Lead Electrocardiogram module with Arduino-based Digital Filters. By using a digital filter. The
contribution of this research is the use of digital filters to remove noise in the electrocardiograph signal. This can be achieved by evaluating
the use of filters on the ECG signal. This research uses Infinite Impulse Filter digital filters such as Butterworth, Chebyshev I, Chebyshev
II, and Elliptic in orders 2, 4, 6, 8, and 10. The research was conducted by providing input from the Fuction Generator on Arduino which
has been applied a digital filter with Frequency 0.5Hz – 100Hz cut-off. Apparatus compared to factory electrocardiograph. The filter
measurement uses a data input of 460. Butterworth filter has the greatest emphasis on the order of 8, the frequency of 0.5Hz produces an
emphasis of -5.74298158 dB and a frequency of 100Hz produces an emphasis of -5.93529424 dB. Chebyshev I filter the greatest emphasis
on the order 6 frequency of 0.5Hz produces a suppression of -3.27104076 dB and on the order of 8 the frequency of 100Hz produces a
suppression of -5.08730424 dB. Chebyshev II filter with the greatest emphasis on the order 8, the frequency of 0.5Hz produces a
suppression of -44.66011104 dB and a frequency of 80Hz produces a suppression of -37.33653957 dB. Elliptic filters with the greatest
emphasis on order 6, the frequency of 0.5Hz produces a suppression of -1.55429354 dB and a frequency of 100Hz on the order of 8
produces a suppression of -2.2849115 dB. The results showed that according to the cut-off frequency is the 6th order Butterworth filter
which is suitable for the application of an electrocardiograph signal filter because it has a bandwidth that suppresses signals outside the
cut-off frequency. The results of this research can be implemented in the 3-Lead EKG module in order to eliminate noise or interference
when tapping the ECG signal.
Keyword—ECG 3 LEAD; Heart Monitoring; Arduino Microcontroller
Abstrak— Pengukuran bio sinyal seperti Elektrokardiograf, memiliki interpretasi noise dari sinyal lain. Noise tersebut
dapat menggangu dalam pengukuran dan membuat pengukuran menjadi tidak akurat. Dalam penelitian terdahulu masih
menggunakan filter analog yang d aplikasikan, maka tujuan dari penelitian ini adalah dibuatnya modul Elektrokardiogram
3-Lead dengan Filter Digital Berbasis Arduino. Dengan mengunakan filter digital. Konstribusi penelitian ini adalah
penggunaan filter digital untuk menghilangkan noise pada sinyal elektrokardiograf. Ini dapat dicapaidengan mengevaluasi
penggunaan filter pada siyal ECG. Penelitian ini menggunakan filter digital Infinite Impulse Filter seperti Butterworth,
Chebyshev I, Chebyshev II, dan Elliptic pada orde 2, 4, 6, 8, dan 10. Penelitian dilakukan dengan memberikan input dari
Fuction Generator pada Arduino yang telah di aplikasikan filter digital dengan Frekuensi cut-off 0.5Hz–100Hz. Alat
dibandingkan dengan Elektrokardiograf pabrik. Pengukuran filter menggunakan input data sebesar 460. Filter
Butterworth penekanan terbesar pada orde 8 frekuensi 0.5Hz menghasilkan penekanan -5,74298158 dB dan frekuensi
100Hz menghasilkan penekanan -5,93529424 dB. Filter Chebyshev I penekanan terbesar pada orde 6 frekuensi 0.5Hz
menghasilkan penekanan -3,27104076 dB dan pada orde 8 frekuensi 100Hz menghasilkan penekanan -5,08730424 dB. Filter
Chebyshev II penekanan terbesar pada orde 8 frekuensi 0.5Hz menghasilkan penekanan -44,66011104 dB dan frekuensi
80Hz menghasilkan penekasan -37,3653957 dB. Filter Elliptic penekanan terbesar pada orde 6 frekuensi 0.5Hz
menghasilkan penekanan -1,55429354 dB dan frekuensi 100Hz pada orde 8 menghasilkan penekanan -2,2849115 dB. Hasil
penelitian menunjukkan bahwa yang sesuai dengan frekuensi cut-off yaitu filter Butterworth orde 6 yang sesuai untuk
pengaplikasian filter sinyal Elektrokardiograf karena memiliki bandwidth yang menekan sinyal diluar frekuensi cut-off.
Hasil penelitian ini dapat diimplementasikan pada modul EKG 3-Lead agar menghilangkan noise atau gangguan pada saat
penyadapan sinyal ECG.
Kata Kunci—ECG 6 LEAD; Heart Monitoring; Arduino Microcontroller
Creator
Syarifudin Abdillah#
, Priyambada C. Nugraha, Dyah Titisari
, Priyambada C. Nugraha, Dyah Titisari
Source
semnas.poltekkesdepkes-sby.ac.id
Date
Surabaya, 28 Nopember 2020
Contributor
PERI IRAWAN
Format
PDF
Language
INDONESIA
Type
TEXT
Files
Citation
Syarifudin Abdillah#
, Priyambada C. Nugraha, Dyah Titisari, “PROSIDING-NASIONAL-KEBIDANAN-2020 POLTEKES KEMENKES CALL FOR PAPER VOL.2 NO 1
Analisis Filter Digital ( Filter IIR pada ECG Pada
Rancang Bangun ECG Lead I, II, Dan III),” Repository Horizon University Indonesia, accessed November 15, 2024, https://repository.horizon.ac.id/items/show/132.
Analisis Filter Digital ( Filter IIR pada ECG Pada
Rancang Bangun ECG Lead I, II, Dan III),” Repository Horizon University Indonesia, accessed November 15, 2024, https://repository.horizon.ac.id/items/show/132.