Analisis Pengaruh Penempatan Sensor Terhadap
Akurasi Pengukuran UV Radiometer Berbasis
Internet Of Things (IOT)
Dublin Core
Title
PROSIDING-NASIONAL-KEBIDANAN-2020 POLTEKES KEMENKES CALL FOR PAPER VOL.2 NO 1
Analisis Pengaruh Penempatan Sensor Terhadap
Akurasi Pengukuran UV Radiometer Berbasis
Internet Of Things (IOT)
Analisis Pengaruh Penempatan Sensor Terhadap
Akurasi Pengukuran UV Radiometer Berbasis
Internet Of Things (IOT)
Subject
Keywords — UV; Angle; Radiometer Device
Description
Abstract— The energy from UV lamps when it hits the surface of human skin can cause skin cancer formation and if it hits the eyes
it can cause cataracts. To find out the Ultraviolet dose from a sterile lamp, measurements using a radiometer are needed and to speed
up the calibration process and avoid continuous exposure to calibration officers. An innovative design of a calibration tool using a UVM
30A sensor that is placed at 16 measurement points to detect UV at an angle of 00
(straight), 100
, 200
, and 300 as well as ESP 32 for
internet network connection through an account on the site of the Internet service provider. Things (IOT), the effect on the sensor when
the angle changes if the angle of view is greater then the value read by the sensor is greater, especially on the sensor that is facing
straight to the lamp an average of 123.75 μW / cm2-153.25 μW / cm2 this is due to greater exposure to UV radiation, the sensor on the
left side of the lamp averaged 127 μW / cm2-156 μW / cm2, while the sensor on the right side of the lamp averaged 110.75 μW / cm2-128.5
μW / cm2, sensors located right and left of the sensor readings are influenced by the lights located next to them. whereas those on 2
lamps show results that tend to be smaller with an average of 33.5 μW / cm2-43.25 μW / cm2, the measurement of 16 sensors on lamp 1
produces the smallest average is sensor 12 and with the most average. large, namely on sensor 15 with a standard deviation of 0.489898 -
0748331. This device only detects UV 00
to 200
, the difference in the maximum value read by the sensor is due to the sensitivity of the
sensor and because the hole in the sensor that detects UV is too small if the sensor is placed incorrectly it will affect the reading value.
Keywords — UV; Angle; Radiometer Device
Abstrak—Energi dari lampu UV apabila mengenai permukaan kulit manusia dapat menimbulkan pembentukan
kanker kulit dan jika mengenai mata dapat menyebabkan katarak. Untuk mengetahui dosis Ultraviolet dari lampu steril
diperlukan pengukuran dengan menggunakan radiometer dan untuk mempercepat proses kalibrasi serta menghindari paparan secara
terus menerus terhadap petugas kalibrasi. Inovasi sebuah rancang bangun suatu alat bantu kalibrasi dengan menggunakan sensor
UVM 30A yang ditempatkan pada 16 titik pengukuran untuk mendeteksi UV pada sudut 00
(lurus), 100
, 200
, dan 300 serta ESP 32
untuk koneksi jaringan internet melalui akun pada situs penyedia layanan Internet Of Things (IOT), pengaruh pada sensor ketika
perubahan sudut jika sudut pengambilan lebih besar maka nilai yang dibaca oleh sensor semakin besar terutama pada sensor yang
berhadapan lurus dengan lampu rata-rata 123,75 μW/ cm2-153,25 μW/ cm2 hal ini dikarenakan paparan radiasi sinar UV lebih besar,
sensor yang berada disamping kiri lampu dengan rata-rata 127 μW/ cm2-156 μW/ cm2, sedangkan sensor yang berada di samping
kanan lampu dengan rata-rata 110,75 μW/ cm2-128,5 μW/ cm2, sensor yang berada dikanan dan kiri lampu hasil pembacaan sensor di
pengaruhi oleh lampu yang berada di sebelahnya. sedangkan yang berada pada 2 lampu menunjukan hasil yang cenderung lebih kecil
dengan rata-rata 33,5 μW/ cm2-43,25 μW/ cm2, pengukuran 16 sensor pada lampu 1 menghasilkan rata-rata terkecil adalah sensor 12
dan dengan rata-rata paling besar yaitu pada sensor 15 dengan standard deviasi 0,489898 – 0748331. Alat ini hanya mendeteksi UV 00
sampai 200
, perbedaan nilai maksimal yang di baca sensor di karenakan sensitifitas sensor dan karena lubang pada sensor
yang mendeteksi uv terlalu kecil jika peletakan sensor tidak tepat maka akan mempengaruhi nilai pembacaan.
Kata Kunci—UV, Sudut, Alat Radiometer.
it can cause cataracts. To find out the Ultraviolet dose from a sterile lamp, measurements using a radiometer are needed and to speed
up the calibration process and avoid continuous exposure to calibration officers. An innovative design of a calibration tool using a UVM
30A sensor that is placed at 16 measurement points to detect UV at an angle of 00
(straight), 100
, 200
, and 300 as well as ESP 32 for
internet network connection through an account on the site of the Internet service provider. Things (IOT), the effect on the sensor when
the angle changes if the angle of view is greater then the value read by the sensor is greater, especially on the sensor that is facing
straight to the lamp an average of 123.75 μW / cm2-153.25 μW / cm2 this is due to greater exposure to UV radiation, the sensor on the
left side of the lamp averaged 127 μW / cm2-156 μW / cm2, while the sensor on the right side of the lamp averaged 110.75 μW / cm2-128.5
μW / cm2, sensors located right and left of the sensor readings are influenced by the lights located next to them. whereas those on 2
lamps show results that tend to be smaller with an average of 33.5 μW / cm2-43.25 μW / cm2, the measurement of 16 sensors on lamp 1
produces the smallest average is sensor 12 and with the most average. large, namely on sensor 15 with a standard deviation of 0.489898 -
0748331. This device only detects UV 00
to 200
, the difference in the maximum value read by the sensor is due to the sensitivity of the
sensor and because the hole in the sensor that detects UV is too small if the sensor is placed incorrectly it will affect the reading value.
Keywords — UV; Angle; Radiometer Device
Abstrak—Energi dari lampu UV apabila mengenai permukaan kulit manusia dapat menimbulkan pembentukan
kanker kulit dan jika mengenai mata dapat menyebabkan katarak. Untuk mengetahui dosis Ultraviolet dari lampu steril
diperlukan pengukuran dengan menggunakan radiometer dan untuk mempercepat proses kalibrasi serta menghindari paparan secara
terus menerus terhadap petugas kalibrasi. Inovasi sebuah rancang bangun suatu alat bantu kalibrasi dengan menggunakan sensor
UVM 30A yang ditempatkan pada 16 titik pengukuran untuk mendeteksi UV pada sudut 00
(lurus), 100
, 200
, dan 300 serta ESP 32
untuk koneksi jaringan internet melalui akun pada situs penyedia layanan Internet Of Things (IOT), pengaruh pada sensor ketika
perubahan sudut jika sudut pengambilan lebih besar maka nilai yang dibaca oleh sensor semakin besar terutama pada sensor yang
berhadapan lurus dengan lampu rata-rata 123,75 μW/ cm2-153,25 μW/ cm2 hal ini dikarenakan paparan radiasi sinar UV lebih besar,
sensor yang berada disamping kiri lampu dengan rata-rata 127 μW/ cm2-156 μW/ cm2, sedangkan sensor yang berada di samping
kanan lampu dengan rata-rata 110,75 μW/ cm2-128,5 μW/ cm2, sensor yang berada dikanan dan kiri lampu hasil pembacaan sensor di
pengaruhi oleh lampu yang berada di sebelahnya. sedangkan yang berada pada 2 lampu menunjukan hasil yang cenderung lebih kecil
dengan rata-rata 33,5 μW/ cm2-43,25 μW/ cm2, pengukuran 16 sensor pada lampu 1 menghasilkan rata-rata terkecil adalah sensor 12
dan dengan rata-rata paling besar yaitu pada sensor 15 dengan standard deviasi 0,489898 – 0748331. Alat ini hanya mendeteksi UV 00
sampai 200
, perbedaan nilai maksimal yang di baca sensor di karenakan sensitifitas sensor dan karena lubang pada sensor
yang mendeteksi uv terlalu kecil jika peletakan sensor tidak tepat maka akan mempengaruhi nilai pembacaan.
Kata Kunci—UV, Sudut, Alat Radiometer.
Creator
Dede Fitra Satria#
, Liliek Soetjiati, Lamidi
, Liliek Soetjiati, Lamidi
Source
semnas.poltekkesdepkes-sby.ac.id
Date
Surabaya, 28 Nopember 2020
Contributor
PERI IRAWAN
Rights
https://drive.google.com/drive/folders/1AernUNLz6pUkVv9tCaGI5xQ2q0IFn2Ua
Format
PDF
Language
INDONESIA
Type
TEXT
Files
Citation
Dede Fitra Satria#
, Liliek Soetjiati, Lamidi, “PROSIDING-NASIONAL-KEBIDANAN-2020 POLTEKES KEMENKES CALL FOR PAPER VOL.2 NO 1
Analisis Pengaruh Penempatan Sensor Terhadap
Akurasi Pengukuran UV Radiometer Berbasis
Internet Of Things (IOT),” Repository Horizon University Indonesia, accessed November 22, 2024, https://repository.horizon.ac.id/items/show/46.
Analisis Pengaruh Penempatan Sensor Terhadap
Akurasi Pengukuran UV Radiometer Berbasis
Internet Of Things (IOT),” Repository Horizon University Indonesia, accessed November 22, 2024, https://repository.horizon.ac.id/items/show/46.