Pengujian Kecepatan Dan Jarak Yang Optimal Pada
Pengiriman Sinyal Dan Detak Jantung Melalui
Bluetooth
Dublin Core
Title
PROSIDING-NASIONAL-KEBIDANAN-2020 POLTEKES KEMENKES CALL FOR PAPER VOL.2 NO 1
Pengujian Kecepatan Dan Jarak Yang Optimal Pada
Pengiriman Sinyal Dan Detak Jantung Melalui
Bluetooth
Pengujian Kecepatan Dan Jarak Yang Optimal Pada
Pengiriman Sinyal Dan Detak Jantung Melalui
Bluetooth
Subject
Keyword—Electrocardiograf; Heart Monitoring; ESP32; Bluetooth
Description
Abstract— Electrocardiograph provides information on the healthy condition of the structure and function of the heart. At this time
many wireless ECGs have been made, including using Bluetooth as a transmission medium. The purpose of this research is to test the
optimal speed and distance for signal transmission and heart rate via Bluetooth in order to obtain accurate results. The design consists of
the ECG Module AD8232, AD8232 Microcontroller IC, and 16x2 character LCD. ECG signal recording on LEAD II. In this study, testing
the performance of Bluetooth transmission and data loss by sending data for 1 second and conducting further analysis. Calibration on
the module is carried out using the ECG Simulator with a simulated heart rate perminute (BPM) and sensitivity settings that have been
determined so that the module design results are in accordance with a standard ECG machine. After evaluating the results data, the error
in the BPM parameter on the module that has been compared with the ECG Recorder is 0.616% and the optimal distance for Bluetooth
transmission is 40 meters without obstruction and with 20 meters with data transmission speed settings of 9600, 19200, 38400, and
115200bps. Testing the optimal speed and distance on this ECG module to prevent a decrease in speed, out of sync, and delay when
transferring data so that signal data transmission and heart rate to the PC must be monitored in real time without delay, high accuracy,
and design. modules can be made inexpensively.
Keyword—Electrocardiograf; Heart Monitoring; ESP32; Bluetooth
Abstrak— Elektrokardiograf memberikan informasi kondisi sehat atau tidaknya struktur dan fungsi jantung. Pada saat ini telah
banyak dibuat sebuah EKG nirkabel jarak jauh (wireless) diantaranya menggunakan Bluetooth sebagai media transmisi. Tujuan
penelitian ini melakukan pengujian kecepatan dan jarak yang optimal untuk pengiriman sinyal dan detak jantung melalui Bluetooth
agar didapatkan hasil akurat. Rancang bangun terdiri dari ECG Module AD8232, IC Mikrokontroller AD8232, dan LCD karakter 16x2.
Perekaman sinyal ECG pada LEAD II. Pada penelitian ini dilakukan pengujian kinerja transmisi Bluetooth dan loss data dengan
mengirim data selama 1 detik dan dilakukan analisis lebih lanjut. Kalibrasi pada modul dilakukan menggunakan ECG Simulator dengan
pengaturan simulasi denyut jantung permenit (BPM) dan sensitivitas yang sudah ditentukan agar hasil rancang bangun modul sesuai
dengan mesin EKG standar. Setelah evaluasi data hasil, kesalahan pada parameter BPM pada modul yang telah dibandingkan dengan
ECG Recorder adalah 0,616% dan jarak optimal transmisi Bluetooth sejauh 40meter tanpa halangan dan dengan halangan 20meter
dengan pengaturan kecepatan pengiriman data 9600, 19200, 38400, dan 115200bps. Pengujian kecepatan dan jarak yang optimal pada
modul ECG ini untuk mencegah adanya penurunan kecepatan, tidak sinkron, dan keterlambatan (delay) ketika transfer data sehingga
pengiriman data sinyal dan detak jantung ke PC yang harus dipantau secara realtime tidak terdapat delay, akurasi tinggi, dan desain
modul dapat dibuat dengan biaya murah.
Kata Kunci—Electrocardiograf; Pemantauan sinyal dan detak jantung ; ESP32; Bluetooth
many wireless ECGs have been made, including using Bluetooth as a transmission medium. The purpose of this research is to test the
optimal speed and distance for signal transmission and heart rate via Bluetooth in order to obtain accurate results. The design consists of
the ECG Module AD8232, AD8232 Microcontroller IC, and 16x2 character LCD. ECG signal recording on LEAD II. In this study, testing
the performance of Bluetooth transmission and data loss by sending data for 1 second and conducting further analysis. Calibration on
the module is carried out using the ECG Simulator with a simulated heart rate perminute (BPM) and sensitivity settings that have been
determined so that the module design results are in accordance with a standard ECG machine. After evaluating the results data, the error
in the BPM parameter on the module that has been compared with the ECG Recorder is 0.616% and the optimal distance for Bluetooth
transmission is 40 meters without obstruction and with 20 meters with data transmission speed settings of 9600, 19200, 38400, and
115200bps. Testing the optimal speed and distance on this ECG module to prevent a decrease in speed, out of sync, and delay when
transferring data so that signal data transmission and heart rate to the PC must be monitored in real time without delay, high accuracy,
and design. modules can be made inexpensively.
Keyword—Electrocardiograf; Heart Monitoring; ESP32; Bluetooth
Abstrak— Elektrokardiograf memberikan informasi kondisi sehat atau tidaknya struktur dan fungsi jantung. Pada saat ini telah
banyak dibuat sebuah EKG nirkabel jarak jauh (wireless) diantaranya menggunakan Bluetooth sebagai media transmisi. Tujuan
penelitian ini melakukan pengujian kecepatan dan jarak yang optimal untuk pengiriman sinyal dan detak jantung melalui Bluetooth
agar didapatkan hasil akurat. Rancang bangun terdiri dari ECG Module AD8232, IC Mikrokontroller AD8232, dan LCD karakter 16x2.
Perekaman sinyal ECG pada LEAD II. Pada penelitian ini dilakukan pengujian kinerja transmisi Bluetooth dan loss data dengan
mengirim data selama 1 detik dan dilakukan analisis lebih lanjut. Kalibrasi pada modul dilakukan menggunakan ECG Simulator dengan
pengaturan simulasi denyut jantung permenit (BPM) dan sensitivitas yang sudah ditentukan agar hasil rancang bangun modul sesuai
dengan mesin EKG standar. Setelah evaluasi data hasil, kesalahan pada parameter BPM pada modul yang telah dibandingkan dengan
ECG Recorder adalah 0,616% dan jarak optimal transmisi Bluetooth sejauh 40meter tanpa halangan dan dengan halangan 20meter
dengan pengaturan kecepatan pengiriman data 9600, 19200, 38400, dan 115200bps. Pengujian kecepatan dan jarak yang optimal pada
modul ECG ini untuk mencegah adanya penurunan kecepatan, tidak sinkron, dan keterlambatan (delay) ketika transfer data sehingga
pengiriman data sinyal dan detak jantung ke PC yang harus dipantau secara realtime tidak terdapat delay, akurasi tinggi, dan desain
modul dapat dibuat dengan biaya murah.
Kata Kunci—Electrocardiograf; Pemantauan sinyal dan detak jantung ; ESP32; Bluetooth
Creator
Sri Wahyuni Dwi Lestari#
, Priyambada Cahya Nugraha, Dyah Titisari
, Priyambada Cahya Nugraha, Dyah Titisari
Source
semnas.poltekkesdepkes-sby.ac.id
Date
Surabaya, 28 Nopember 2020
Contributor
PERI IRAWAN
Rights
https://drive.google.com/drive/folders/1AernUNLz6pUkVv9tCaGI5xQ2q0IFn2Ua
Format
PDF
Language
INDONESIA
Type
TEXT
Files
Citation
Sri Wahyuni Dwi Lestari#
, Priyambada Cahya Nugraha, Dyah Titisari, “PROSIDING-NASIONAL-KEBIDANAN-2020 POLTEKES KEMENKES CALL FOR PAPER VOL.2 NO 1
Pengujian Kecepatan Dan Jarak Yang Optimal Pada
Pengiriman Sinyal Dan Detak Jantung Melalui
Bluetooth,” Repository Horizon University Indonesia, accessed November 15, 2024, https://repository.horizon.ac.id/items/show/50.
Pengujian Kecepatan Dan Jarak Yang Optimal Pada
Pengiriman Sinyal Dan Detak Jantung Melalui
Bluetooth,” Repository Horizon University Indonesia, accessed November 15, 2024, https://repository.horizon.ac.id/items/show/50.